Fluke 725 User Manual
Page 34

725
Users Manual
24
Table 6. RTD Types Accepted
RTD Type
Ice Point (R
0
)
Material
α
Range (
°
C)
Pt100 (3926)
100
Ω
Platinum
0.003926
Ω
/
°
C
-200 to 630
Pt100 (385)
100
Ω
Platinum
0.00385
Ω
/
°
C
-200 to 800
Ni120 (672)
120
Ω
Nickel
0.00672
Ω
/
°
C
-80 to 260
Pt200 (385)
200
Ω
Platinum
0.00385
Ω
/
°
C
-200 to 630
Pt500 (385)
500
Ω
Platinum
0.00385
Ω
/
°
C
-200 to 630
Pt1000 (385)
1000
Ω
Platinum
0.00385
Ω
/
°
C
-200 to 630
Pt100 (3916)
100
Ω
Platinum
0.003916
Ω
/
°
C
-200 to 630
The Pt100 commonly used in U.S. industrial applications is Pt100 (3916),
α
= 0.003916
Ω
/
°
C. (Also designated as JIS
curve.) The IEC standard RTD is the Pt100 (385),
α
= 0.00385
Ω
/
°
C.