Mtm module — 23, 4848 reactor controller – Parr Instrument 4848 User Manual
Page 24

4848 Reactor Controller
P a r r I n s t r u m e n t C o m p a n y
24
MTM Module
Press return key and release
Operation
Mode
Select
type/value Comment
r-S
Run
Run/Stop
SP
1
Decimal point position
LoC
OFF
Lock mode (lock all keys or
only up/down arrow able to
use)
Press and hold down “SET” for 5-sec
Operation
Mode
Select
type/value Comment
InPt
v10
Input type (v5 = 0-5V, v10 =
0-10V, nA0 = 0-20mA, nA4 =
4-20mA, nV =0-50mV)
tP-H
To be deter-
mined**
Upper-limit range
tP-L
0
Lower-limit temperature
range
CTRL
ON/OFF
Control mode (ON/OFF,
MANUAL, PID and PID
PROG)
S-HC
Heat
Heat/Cool control
ALA1
0
Alarm mode
SALA
OFF
System alarm
CoSH
ON
Communication write func-
tion that able to use set
point from software
C-Sl
RTU
Format type
C-no
4
Communication address
bPS
9600
Communication baud rate
Len
8
Data length setting
PrtY
Even
Parity bit setting
StoP
1
Stop bit setting
** For 90VDC motor, 115VAC, tP-H = 142.9
For 180VDC motor, 230VAC, tP-H = 120.4
For 90VDC motor, 230VAC, tP-H = 153.2
Press “SET” and release
Operation
Mode
Select
type/value
Comment
HtS
0
Hysteresis
TPoF
0
PV Offset
CrHI
0
Analog output high
CrLo
0
Analog output low