beautypg.com

3 brief operating instructions – Martin Christ RVC 2-18 CDplus User Manual

Page 75

background image

RVC 2-18 CD& HCL

11 Appendix

Version 04/2012, Rev. 1.4 of 25/11/2013 • sb

75

Translation of the original operating manual

Pos: 202 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 18 CD plus_ 2-1 8 CDplus _HCl/11 0 Anh ang/ 110 -00 41 Kurzb edie nun gsanl eitun g HCl @ 12\m od_ 133 542 656 832 9_68 .docx @ 6 405 8 @ 2 @ 1

11.3 Brief operating instructions

1. Switch the system on via the mains power switch on the right side of the

unit.

2. Switch the vacuum pump and/or the cooling trap on.
3. Close the lid.
4. Starting the warm-up phase:

• Enter the set values for the warm-up phase under "menu"

"options" "timer warm-up" and "temperature warm-up". Confirm
your entries and quit the menu.

• Press the left-hand function key "mode".
• Select the "warm-up" menu and confirm.
• Open the lid when the warm-up phase has ended.
• Install the rotors and close the lid.

5. Changing the set values for time, temperature, vacuum and safety

pressure:
• Select "menu" with the right-hand function key.
• Select "change set values for manual mode" from the main menu.
• Select the set value with the aid of the up/down keys and confirm

with the right-hand function key "edit".

• Change the set value with the aid of the up/down keys and confirm

with the right-hand function key "ok".

• Quit the menu with the left-hand function key.

6. Starting the process "evaporation manual":

• Open the "mode" menu with the left-hand function key.
• Select "evaporation manual" using the up/down keys.
• Start the evaporation process by pressing the right-hand function

key "enter".

The start of the evaporation process starts the following automatic
process run:

• The rotor speed increases (direction of rotation: anti-clockwise)
• At 70% of the selected maximum speed: the aeration valve closes

automatically (to avoid boiling retardation).

• At 80% of the selected maximum speed: the stop valve opens the

connection to the vacuum pump

• Further speed increase until the preselected maximum rotor speed

is reached.

7. Interrupting an evaporation process:

• Press the left-hand function key "mode".
• Select "stop evaporation" and confirm with the right-hand function

key.

Without an interruption, the evaporation process ends after the
preselected time. The following process runs automatically:

• At maximum speed: the stop valve closes the connection to the

vacuum pump, the aeration valve opens; the rotor speed is
maintained for 30 sec (aeration phase, to avoid boiling retardation).

• The rotor decelerates until it is at a standstill. After the standstill of

the rotor the lid can be opened in order to remove the rotors.