7 Malfunctions and error correction
Version 04/2012, Rev. 1.4 of 25/11/2013 • sb
47
Translation of the original operating manual
Pos: 141 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 Stör ung en u nd Fehl ersuc he === == == === == == === == == === == = @ 8\m od_ 1309 939 131 742 _68. docx @ 47 129 @ 1 @ 1
7 Malfunctions and error correction
Pos: 142 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10 Allg emei ne Stö run gen @ 8\ mod _13 099 413 440 43_ 68.d ocx @ 472 24 @ 2 @ 1
7.1 General malfunctions
Malfunctions are displayed in the menu "Process & equipment info" (see
chapter 6.5.3.2 - "Process and equipment information"). A sound signal
sounds when an error message is generated.
• Eliminate the source of the problem (see the following chapters).
• Acknowledge the error messages by pressing the right-hand function
key.
Pos: 143 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 144 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 18 CD plus_ 2-1 8 CDplus _HCl/07 0 Stö run gen und Fe hlers uche /070 -00 10 -001 0 Fehl er meldu nge n RVC 2- 18_ 2-1 8 HCl @ 12\ mod _13 359 3986 448 7_6 8.do cx @ 6414 1 @ 3 @ 1
7.2 Error messages
Error
Possible reason
Correction
Lid cannot be opened
• The rotor chamber is not
aerated completely
• Aerate the rotor chamber
completely
Lid contacts error
• The system has detected an
open lid during the evaporation.
The rotor will be stopped
immediately. The lid must be
closed while the rotor is moving.
• Check the lid contacts. If
necessary, contact the after-
sales service (see chapter 7.3 -
"Service contact")
Rotor overtemperature
• The maximum permissible rotor
temperature has been
exceeded
• If the system does not cool
down, disconnect it immediately
from the power supply.
• Contact the after-sales service
(see chapter 7.3 - "Service
contact")
Chamber overtemperature
• The maximum permissible
chamber temperature has been
exceeded.
• If the system does not cool
down, disconnect it immediately
from the power supply.
• Contact the after-sales-service
(see chapter 7.3 - "Service
contact").
Permissible imbalance exceeded
• The rotor will be stopped
immediately, the process will
also be stopped.
– improper loading of the
rotor
– glass breakage during the
evaporation
– uneven evaporation
• Eliminate the imbalance and
restart the unit. If the error is still
present, contact the after sales-
service (see chapter 7.3 -
"Service contact")
VSK3000 calibration error
• Incorrect calibration of the
vacuum sensor VSK3000
• The sensor must be recalibrated
(see the separate operating
manual).