Routine maintenance – Labconco Protector Controlled Atmosphere Glove Boxes 5080102 User Manual
Page 31

ROUTINE MAINTENANCE
32
• Stretch the new glove’s beaded cuff over the old glove cuff and into the groove nearest the
window.
• To remove the old glove, grasp the new glove surface and manipulate the old glove bead
free from the outer groove.
• With the other glove hand inside the box, pull the old glove into the glove box.
• Install the O-ring over the new glove onto the outer groove.
Fluorescent Lamp Replacement
• Turn the light switch off and unplug the lamp cord from the receptacle located on the
outside top of the cabinet.
• From the top of the lamp housing, remove the two Phillips-type screws attaching the lamp
assembly to the top edge of the window frame.
• Remove the lamp assembly from the unit.
• The lamp tube can then be easily removed and replaced. Replacement lamp part number
12779, fluorescent lamp #F30T12/CW/RS may be purchased from most local lighting
supply stores.
• Re-install the lamp assembly and attach with the two Phillips-type screws. Plug in the
power cord and switch on the lamp. Refer to the illustration below.
Figure 17
- Protector Controlled Atmosphere Glove Boxes 5070102 Protector Controlled Atmosphere Glove Boxes 5060102 Protector Controlled Atmosphere Glove Boxes 5080100 Protector Controlled Atmosphere Glove Boxes 5070100 Protector Controlled Atmosphere Glove Boxes 5060100 Protector Controlled Atmosphere Glove Boxes 5080002 Protector Controlled Atmosphere Glove Boxes 5070002 Protector Controlled Atmosphere Glove Boxes 5060002 Protector Controlled Atmosphere Glove Boxes 5080000 Protector Controlled Atmosphere Glove Boxes 5070000 Protector Controlled Atmosphere Glove Boxes 5060000