beautypg.com

Labconco XVS Ventilation Stations 4865060 User Manual

Page 49

background image

Chapter 7: Accessorizing your XPert or XVS Enclosure

Product Service 1-800-522-7658

45

Catalog #

HP

Description

CFM@RPM Ranges at Static Pressure (in H

2

0)

Std. EP

.25" .38" .50" .62" .75" .88" 1.00"

7068200

1/4

Remote
Blower, 115V,
60 Hz, 4.4 A

540@800

410@800 350@870 390@970 430@1060

7068300

1/4

Explosion
Proof Blower
115V, 60 Hz,
4.5 A

720@950 760@1050 710@1060 620@1060

7068400

1/3

Remote
Blower, 115V,
60 Hz, 6.1 A

760@1050 710@1060 620@1060 430@1060 460@1150 500@1220

7068500 1/3

Explosion
Proof Blower,
115V, 60 Hz,
6.4 A

850@1130 825@1160 790@1190 770@1220 740@1260 640@1260














6. Exhaust Dampers

Exhaust dampers allow an adjustment required to maintain proper airflow
for roof-mounted blowers or house exhaust systems.

Catalog #

Description

3924000

6" Epoxy Coated Steel In-Line adjustable damper

4724200

6" PVC In-line adjustable damper

5983400

10" PVC In-line adjustable damper

7. FilterMate Portable Exhausters and Filters

For filtered exhaust, Labconco offers FilterMate Portable Exhausters
capable of exhausting up to 280 cfm of HEPA filtered air or up to 220
cfm of combination HEPA/Carbon filtered air when connected to the
enclosure. For additional exhaust volume, multiple FilterMates with
balanced airflow can be used.

This manual is related to the following products: