7

HI 93710 - pH & Chlorine

HI 93710 - pH & Chlorine

HI 93710 - pH & Chlorine

HI 93710 - pH & Chlorine

HI 93710 - pH & Chlorine

The HI 93710 meter measures both pH and chlorine (Cl

2

, free and

total) content in water and wastewater in the following ranges:
pH

5.9 to 8.5 pH units

Free Chlorine

0.00 to 2.50 mg/L (ppm)

Total Chlorine

0.00 to 3.50 mg/L (ppm).

SPECIFICATIONS

Range pH 5.9 to 8.5

Free Cl

2

0.00 to 2.50 mg/L

Total Cl

2

0.00 to 3.50 mg/L

Resolution

0.1 pH/0.01 mg/L Cl

2

Accuracy

±0.1 pH

±0.03 mg/L ±3% of Cl

2

reading

Typical EMC

±0.2 pH

Deviation

±0.02 mg/L Cl

2

Light Source

Light Emitting Diode @ 555 nm

Method

Adaptation of the EPA recommended DPD method

330.5 for chlorine analysis. The reaction with

reagents causes a pink tint in the sample.
For pH, Phenol red method. The reaction with

reagents causes a red tint in the sample.

REQUIRED REAGENTS

Code

Unit

Description

Quantity

HI 93710-0

pH

Phenol red

5 drops

HI 93701-0

Free Cl

2

DPD

1 packet

HI 93710

pH/Chlorine

This manual is related to the following products: