• Fill one graduated mixing cylinder up to the

25 mL mark with deionized water.

• Add the content of one packet of HI 93746 TPTZ reagent, close

the cylinder and agitate for 30 seconds. This is the blank.

• Fill a cuvet with 10 mL of the blank up

to the mark and let it sit for 3 minutes.

• Replace the cap, insert the blank into

the holder and make sure that the notch
on the cap is positioned securely into
the groove.

• Press ZERO and "SIP" will appear on the display.

• Wait for a few seconds and the display will

show "-0.0-". Now the meter is zeroed and
ready for measurement.

• Fill one graduated mixing cylinder up to the

25 mL mark with the sample.

• Add the content of one packet of HI 93746 TPTZ reagent, close

the cylinder and agitate for 30 seconds.