Cyanuric acid – Hanna Instruments HI 93710 User Manual

Page 38

background image

38

• Fill the cuvet up to the mark with 10 mL of

sample and replace the cap.

• Place the cuvet into the holder and

ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO and "SIP" will appear on the display.

• Wait for a few seconds and the display will

show "0". Now the meter is zeroed and ready
for measurement.

• Fill a beaker up to the

25 mL mark with the
sample, add the con-
tent of one packet of
HI 93722 reagent and
stir gently to mix.

• Fill a second cuvet with 10 mL of the reacted

sample up to the mark. Replace the cap.

• Reinsert the cuvet into the instrument.

• Wait for 45 seconds and press READ. "SIP" will appear during

measurement.

• The instrument directly displays concentration in mg/L of cyanuric

acid on the Liquid Crystal Display.

ZERO

CYANURIC ACID

CYANURIC ACID

CYANURIC ACID

CYANURIC ACID

CYANURIC ACID MEASUREMENT

MEASUREMENT

MEASUREMENT

MEASUREMENT

MEASUREMENT

Cyanuric Acid

12

3

6

9

45"

10 mL

10 mL