beautypg.com

Hanna Instruments HI 93710 User Manual

Page 15

background image

15

The HI 93742 meter measures the iron and manganese content in
water, wastewater and seawater (iron only) in the following ranges:
Iron

0 to 400 µg/L (ppb)

Manganese 0 to 300 µg/L (ppb).

SPECIFICATIONS

Range

Fe 0 to 400 µg/L

Mn 0 to 300 µg/L

Resolution

1 µg/L

Accuracy

Fe ±10 µg/L ±8% of reading

Mn ±2 µg/L ±3% of reading

Typical EMC

±1 µg/L

Deviation
Light Source

Light Emitting Diode @ 555 nm

Method

Adaptation of the TPTZ method for iron and

PAN method for manganese. The reaction be-

tween iron or manganese and the reagents

respectively causes a blue or an orange tint in

the sample.

REQUIRED REAGENT

Code

Description

Quantity

Iron

HI 93746-0

TPTZ reagent

2 packets

HI 93742

Iron LR/Manganese LR

HI 93742 -

HI 93742 -

HI 93742 -

HI 93742 -

HI 93742 - Iron Low Range and

Iron Low Range and

Iron Low Range and

Iron Low Range and

Iron Low Range and

Manganese Low Range

Manganese Low Range

Manganese Low Range

Manganese Low Range

Manganese Low Range

This manual is related to the following products: