beautypg.com

Hanna Instruments HI 93710 User Manual

Page 24

background image

24

HI 93745
Chlorine/pH/Iron LR/Hardness

Light Source

Light Emitting Diode @ 555 nm

Method

For pH, Phenol red method. The reaction be-
tween pH and the reagent causes a red tint in
the sample.
For Chlorine, adaptation of the EPA recom-
mended DPD method 330.5. The reaction
between chlorine and the reagent causes a pink
tint in the sample.
For Iron, Adaptation of the TPTZ Method. The
reaction between iron and the reagent causes a
blue tint in the sample.
For Hardness, adaptation of the Standard Meth-
ods for the Examination of Water and
Wastewater, 18

t h

edition,

Calmagite/colorimet-

ric method. The reaction between hardness and
reagents causes a maroon tint in the sample.

REQUIRED REAGENTS

Code

Unit

Description

Quantity

HI 93710-0

pH

Phenol red

5 drops

HI 93701-0

Free Cl

2

DPD

1 packet

HI 93711-0

Total Cl

2

DPD

1 packet

HI 93746-0

Iron LR

TPTZ Reagent

2 packets

HI 93719A-0 Hardness

Ca & Mg indicator 0.5 mL

HI 93719B-0 Hardness

Alkali solution

0.5 mL

HI 93719C-0 Hardness

EDTA solution

1 drop

HI 93719D-0 Hardness

EGTA solution

1 drop

Liquid version (chlorine):

HI 93701A-F Free Cl

2

DPD1 indicator

3 drops

HI 93701B-F Free Cl

2

DPD1 buffer

3 drops

HI 93701A-T Total Cl

2

DPD1 indicator

3 drops

HI 93701B-T Total Cl

2

DPD1 buffer

3 drops

HI 93701-C Total Cl

2

DPD3 solution

1 drop

REAGENT SETS

HI 93710-01 Reagents for 100 pH tests
HI 93710-03 Reagents for 300 pH tests
HI 93701-01 Reagents for 100 free chlorine tests
HI 93701-03 Reagents for 300 free chlorine tests
HI 93711-01 Reagents for 100 total chlorine tests
HI 93711-03 Reagents for 300 total chlorine tests
HI 93701-F Reagents for 300 free chlorine tests (liquid version)

This manual is related to the following products: