• Fill a cuvet with 10 mL of the reacted

sample up to the mark and replace the
cap. This is the sample.

• Insert the sample into the instrument.

• Wait for 3 minutes and then press READ. "SIP" will appear

during measurement.

• The instrument directly displays concentration in µg/L of iron on

the Liquid Crystal Display.

Note

Note

Note

Note

Note: For better accuracy wash glassware with HCl 6N.

INTERFERENCES

Interference may be caused by:
Cadmium above 4.0 mg/L
Chromium

3+

above 0.25 mg/L

Chromium

6+

above 1.2 mg/L

Cobalt above 0.05 mg/L
Copper above 0.6 mg/L
Cyanide above 2.8 mg/L
Manganese above 50.0 mg/L
Mercury above 0.4 mg/L
Molybdenum above 4.0 mg/L
Nickel above 1.0 mg/L
Nitrite ion above 0.8 mg/L
Sample pH should be between 3 and 4 to avoid developed color to
fade or turbidity.

Iron LR

10 mL