Hanna Instruments HI 93710 User Manual

Page 40

background image

40

• Nothing is added to the third cuvet.

• Place the ZERO1 sample into the holder

and ensure that the notch on the cap
is positioned securely into the groove.

• Press ZERO; "SIP" will appear during measurement, then the

instrument will display "ZER2".

• Remove the ZERO1 sample and insert

the ZERO2 sample into the instru-
ment.

• Press ZERO again; "SIP" will appear during measurement, then

the instrument will display the level of Magnesium hardness in
ppm CaCO

3

(together with "n").

• Remove the ZERO2 sample and insert

the third cuvet into the instrument.

• Press READ; "SIP" will appear during measurement, then the

instrument will display the concentration of Calcium in ppm CaCO

3

(together with "C").

• Press READ again; "SIP" will appear during measurement and

the instrument will display the total hardness concentration in
ppm CaCO

3

(together with "t").

ZERO

ZERO2

ZERO

3rd CUVET

Hardness

ZERO1