Hanna Instruments HI 93710 User Manual

Page 17

background image

17

H

I9

3

7

4

8

A

B

C

• Add the content of one packet of

HI 93748A ascorbic acid to each
cuvet, replace the caps and shake
gently.

• Add 0.2 mL of the

HI 93748B alkaline-
cyanide reagent
solution to each cu-
vet, replace the
caps and shake gen-
tly.

• Add 1 mL of the HI 93748C

0.1% PAN indicator solution
to each cuvet, replace the
caps and shake gently.

• Place the cuvet with the reacted deion-

ized water (blank) into the holder
and ensure that the notch on the cap
is positioned securely into the groove.

• Wait for 2 minutes and press ZERO. "SIP" will appear during

zeroing.

• Wait for a few seconds and the display will

show "-0.0-". Now the meter is zeroed and
ready for measurement.

• Insert the second cuvet with the reacted

sample into the instrument.

ZERO

12

3

6

9

2’

HI 93742

Iron LR/Manganese LR