beautypg.com

Setup mode – Hanna Instruments HI 700 Series User Manual

Page 7

background image

13

12

• Then confirm the displayed digit with

and move to the next one.

• When the whole password has been

inserted, press CFM to confirm it.

Note

The default password is set at “0000”.

• The LCD will display “SET” on the

upper part and “c.00” on the lower,
allowing the user to edit setup param-
eters (see table below).

• Enter the code of the parameter you want to set, using the

arrow keys as per the password procedure above (e.g. 41).

• Confirm the code by pressing CFM and the default or the

previously memorized value will be displayed with the first
digit blinking.

Note

When the password is not inserted or a wrong password is
confirmed, the display will only show the previously memo-
rized value, without blinking (read only mode). In this case,
the value cannot be set. Press LCD and start again.

• Enter the desired value using the arrow keys and then press

CFM.

SETUP MODE

HI 700 and HI 710 offer a multitude of possibilities from
ON/OFF or PID dosage to analog recorder output and from
alarm to selftest features.
The Setup Mode allows the user to set all needed character-
istics of the meter.

The setup mode is entered by pressing SETUP
and entering the password when the device is
in idle or control mode.
Generally speaking, if the password is not inserted the user
can only view the setup parameters (except for password)
without modifying them (and the device remains in control
mode). An exception is certain setup items, or flags, which
can activate special tasks when set and confirmed.
Each setup parameter (or setup item) is assigned a two-digit
setup code which is entered and displayed on the secondary
LCD.

The setup codes can be selected after password and CFM
are pressed. When CFM is pressed, the current setup item is
saved on EEPROM and the following item is displayed.
Whenever LCD is pressed, the de-
vice reverts back to control mode.
The same is true when CFM is
pressed on the last setup item.

The possible transitions in setup mode are the following:

ENTERING THE PASSWORD

• Press SETUP to enter the setup mode. The LCD will display

“0000” on the upper part and “PAS” on the lower. The first
digit of the upper part of the LCD will blink.

• Enter the first value of the pass-

word by the or keys.

This manual is related to the following products: