Installation, Specifications – Hanna Instruments HI 700 Series User Manual
Page 5

9
8
INSTALLATION
HI 700 and HI 710
series offer a multitude
of possibilities, from
single and dual set-
points to ON/OFF or
PID dosage, isolated
outputs with user-select-
able zoom, bi-directional
RS485, recorder outputs
in mAmps and Volts.
Use the 3-wire Pt 100
temperature sensor to
compensate for the
cable resistance and
have a precise auto-
matic temperature
compensation of the
measurements in long
distance applications.
See the diagram for a
recommended installa-
tion.
SPECIFICATIONS
Ranges
0.0 to 199.9
μS, 0 to 1999 μS
0.00 to 19.99 mS, 0.0 to 199.9 mS
0.0 to 100.0 ppm, 0 to 1000 ppm (HI 710* only)
0.00 to 10.00 ppt, 0.0 to 100.0 ppt (HI 710* only)
-10.0 to 100.0 °C
Resolution
0.1 μS, 1 μS, 0.01 mS, 0.1 mS
0.1 ppm, 1 ppm (HI 710 models only)
0.01 ppm, 0.1 ppm (HI 710 models only)
0.1 °C
Accuracy
±0.5 % F.S. (EC and TDS)
(@20°C/68°F)
±0.5°C (0 to 70°C), ±1°C (outside)
Typical EMC Deviation
±2 % F.S. (EC and TDS) / ±0.5 °C
EC Calibration
Automatic or manual at 1 point
Temperature Compensation
Automatic or manual, -10 to 100°C with adjustable
temperature coefficient from 0.00 to 10.00%/°C
TDS Factor (HI 710 only)
Adjustable from 0.00 to 1.00
Outputs
Analog: isolated 0-1 mA, 0-20 mA and 4-20 mA;
0-5 Vdc, 1-5 Vdc and 0-10 Vdc; or digital: RS485
Analog Input
4-20 mA
Setpoint Relay
2 contact outputs SPDT, 5A-250 Vac, 5A-30 Vdc
(resistive load). Fuse protected: 5A, 250V FUSE
Alarm Relay
Contact output SPDT, 5A-250 Vac, 5A-30 Vdc
(resistive load). Fuse protected: 5A, 250V FUSE
Power Supply
230 ±10% Vac or 115 ±10% Vac, 50/60 Hz
Power Consumption
15 VA
Over Current Protection
400 mA 250V FAST FUSE
Environment
0 to 50°C (32 to 122°F); RH max 95%
Installation Category
II
Enclosure
Single case ½ DIN
Dimensions
Panel cutout: 140 x 140 mm
Instrument: 144 x 144 x 170 mm
Weight
Approximately 1.6 kg. (3.5 lb.)
* Note: actual TDS range for HI 710 models depends on TDS factor set.