value shown on the secondary display (e.g. 4).

• Wait for approximately 30 seconds (until the reading of the

calibrator is stable).

• Press CFM to confirm. The meter will switch to the second

calibration point. Repeat the above procedure.

• After the desired readings are obtained, press CFM and

the meter will skip back to normal operational mode.

Note

When adjusting values using the or keys it is important
to allow for sufficient response time (up to 30 seconds)

The table below lists the values of output codes along with
the calibration point values (which are the analog output
minimum and the analog output maximum) as indicated on
the display.

The secondary display indicates the current calibration point
value, while primary display indicates the current calibration
type.

OUTPUT

CALIBRATION

CALIBRATION

CALIBRATION

TYPE

CODE

POINT 1

POINT 2

0-1 mA

0

0 mA

1 mA

0-20 mA

1

0 mA

20 mA

4-20 mA

2

4 mA

20 mA

0-5 Vdc

3

0 Vdc

5 Vdc

1-5 Vdc

4

1 Vdc

5 Vdc

0-10 Vdc

5

0 Vdc

10 Vdc

ANALOG OUTPUT CALIBRATION

In the meters where the analog output is available, this fea-
ture is factory calibrated through software. The user may also
perform these calibration procedures.

IMPORTANT

It is recommended to perform the out-
put calibration at least once a year.
Calibration should only be performed
after 10 minutes from power up.

• With a multimeter or an HI 931002 connect the common

port to the ground output and the second port to the cur-
rent output or voltage output (depending on which
parameter is being calibrated).

• Press and hold in sequence CFM first, then and then

CAL to enter the Analog Output Calibration mode.

• Execute the password procedure.

• The primary display will show the current selected param-

eter blinking. Use the to select the code (0-5 see chart
below) for the desired parameter displayed on the second-
ary display (e.g. 4-20 mA).

• Press CFM to confirm the selected parameter that will stop

blinking on the primary display. The secondary display shows
the HI 931002 or multimeter input value as lower limit of
the interval.

• Use the or to make the

HI 931002 or multimeter out-
put correspond with the meter’s