ACCESSORIES

CONDUCTIVITY CALIBRATION SOLUTIONS

HI 7030L

12880 μS/cm calibration solution, 500 mL

HI 7030M

12880 μS/cm calibration solution, 230 mL

HI 7031L

1413 μS/cm calibration solution, 500 mL

HI 7031M

1413 μS/cm calibration solution, 230 mL

HI 7033L

84 μS/cm calibration solution, 500 mL

HI 7033M

84 μS/cm calibration solution, 230 mL

HI 7034L

80000 μS/cm calibration solution, 500 mL

HI 7034M

80000 μS/cm calibration solution, 230 mL

HI 7035L

111800 μS/cm calibration solution, 500mL

HI 7035M

111800 μS/cm calibration solution, 230 mL

HI 7039L

5000 μS/cm calibration solution, 500 mL

HI 7039M

5000 μS/cm calibration solution, 230 mL

OTHER ACCESSORIES

HI 731326

Calibration screwdrivers (20 pcs)

HI 779/P

6-wire cable (100 m/330' roll)

HI 8733

Portable conductivity meter with automatic tem-
perature compensation

PROBE MAINTENANCE & CLEANING

The probe can be compensated for normal contamination by a process of
re-calibration.
It is recommended to remove the probe from its installation regularly for
maintenance.

For HI 7635 only:
Deposits on the conductivity probe can be removed by immersing the probe
in 0.1 N hydrochloric acid for about 30 minutes. Heavier deposits may
demand longer immersion periods.
Rinse the probe thoroughly with water prior to the reinstallation.
On reinstalling, check the seals carefully to ensure that a leak-free
connection is obtained.

For HI 7638 only:
Rinse the probe with tap water. If a more thorough cleaning is desired,
unscrew the plastic sleeve and clean the sensors with a nonabrasive cloth
or alcohol.
After cleaning the probe, re-calibrate the instrument. If the instrument
will not calibrate with a clean probe, you must replace the probe.

Note: Always re-calibrate the meter when attaching a new probe.