beautypg.com

Hanna Instruments HI 83414 User Manual

Page 41

background image

41

LAMP REPLACEMENT

The instrument tungsten lamp has a life longer than 100,000 measurements. In case of lamp
failure, the defective lamp can be easily replaced. When the lamp is broken, the instrument displays
a lamp error message.

To replace the lamp follow the next steps:
• Power off the instrument and take out the

power cord.

• Remove the fixing screw of the lamp lid.
• Remove the lamp lid.
• Unscrew the lamp leads from connector.
• Unlock the lamp and extract it by pulling it out

from the lamp holder.

• Place the new lamp in the right position and

push it until is securely locked.

• Insert the lamp leads into the connector and

tighten them using a screwdriver.

• Return the lamp lid and secure with the screw.
Warning: After lamp replacement the meter has

to be recalibrated.

FUSE REPLACEMENT

To change the fuse follow next steps:
•Disconnect the power cord from the rear panel of the instrument.
•Pull out the fuse holder located near the power cord connector.
•Replace the fuse with a similar one.
•Push the fuse holder with the fuse in the appropriate place.