Hanna Instruments HI 83414 User Manual

Page 15

background image

15

• Insert the cuvette into the instrument and press

the

Read key. Record the reading.

• Open the instrument lid, slightly rotate the

cuvette and take a new reading.

• Repeat the last step until you read the lowest

NTU value.

• Alternatively, keep the

Read key pressed to

make continuous readings. After the first value
is displayed, open the lid and start rotating the
cuvette until the lowest NTU value is displayed.

• Mark this position on the thicker white band on

the top of the cuvette with a water resistant
pencil.

• Always use this position to align it with the

sign on the instrument top.

In order to match multiple cuvettes follow the next
steps:
• Fill some cuvettes with high quality water

(<0.1NTU) up to the 10 mL mark.

• Clean and oil the cuvettes as described before.

MATCHING MULTIPLE CUVETTES

Precise measurements require the use of a single cuvette. If it is not possible, cuvette selection
and matching must be performed before taking measurements.