• Turn the instrument ON.

• Insert the first cuvette into the instrument and

press the

Read key. Record the reading.

• Mark this position on the thicker white band on

the top of the cuvette with a water resistant
pencil.

• Insert the second cuvette into the instrument

and take a reading.

• Open the instrument lid, slightly rotate the

cuvette and take a new reading.

• Repeat the last step for the second cuvette until the reading is within 0.01 NTU of the value

obtained for the first cuvette.

16