To calibrate one range, follow next steps:
• Turn the instrument on by toggling ON/OFF.

When main screen appears, the instrument is
ready.

• Make sure that the correct range is selected

free or total chlorine. If necessary, press

RANGE

to select the correct range.

• Press

CAL CHECK and then press the Cal key

to start calibration procedure.

• Place the zero cuvette (A) in the instrument

with the mark on the cuvette aligned with the
mark on the instrument top.

• Press the

Zero key. The dashes and the lamp

icon will blink on the LCD, depending on the
measurement phase. If the zeroing procedure
was successful, the display will show “

-0.0-”.

• Remove the cuvette.

• Place the CAL CHECK™ Standard Cuvette B

into the holder. Make sure that the mark on
the glass is aligned with the mark on the
instrument top.

• Press the

Read key. The LCD wil display dashes

and the lamp icon blinking, depending on the
measurement phase. After a few seconds the
display will briefly show the calibration
standard value, store the calibration without
warning and return to the measurement screen.

Use the provided standard solution to calibrate the free or total chlorine range of the

HI 83414. Do

not calibrate the instrument with standard solutions other than Hanna CAL CHECK™ Standards,
otherwise erroneous results will be obtained. For accurate calibration please perform test at room
temperature, 18 to 25 ºC (64.5 to 77.0 ºF).

CALIBRATION PROCEDURE FOR FREE/TOTAL CHLORINE