• Place the cuvette into the instrument. Align the

mark on the cuvette with the sign on the
instrument top and close the lid.

• Press the

Zero key. The dashes and the lamp

icon will blink on the LCD, depending on the
measurement phase. If the zeroing procedure
was successful, the display will show “

-0.0-”.

SINGLE SAMPLE READ

• Remove the cuvette from the instrument.

• Remove the cap.

• Add the content of one packet of the specific

test reagent, for:

Free Chlorine

Total Chlorine

1 packet of or 1 packet of
HI 93701-0

HI 93711-0

• Replace the cap and shake gently for 20 seconds

(or 2 minutes in case of seawater analysis).

• Replace the cuvette into the holder and ensure

that the mark on the glass matches the mark on
the instrument top.

• Close the lid.