UT OF CAL RANGE FUNCTION

The instrument has a mechanism to prevent taking
measurements in a range where the calibration
does not assure the best results. The message

Out of Calibration Range” is displayed on the

message line in the following situations:
• When the first calibration point is over 0.15 NTU and the reading is under 10 NTU.
• When two point calibration was performed and the reading value is over 40 NTU.
• When a three point calibration was performed and the reading is over 150% of the the third point value.
• When a four point calibration was performed and the reading is over 200% of the the fourth point value.

To take colorimetric measurements follow next
steps:
• Turn the instrument on by pressing ON/OFF.
• Be sure that the correct range is selected

(displayed on the down-right side of the LCD).
Only the

Zero key is available, indicating that

a zero measurement should be done.

ZEROING THE INSTRUMENT

• Fill a clean, dry cuvette with 10 mL of sample,

up to the mark, taking care to handle the cuvette
by the top.

• Replace the cap.

• Wipe the cuvette thoroughly with a lint-free

cloth to remove any fingerprints, dirt or water
spots.

MEASUREMENT PROCEDURE FOR FREE/TOTAL CHLORINE

When taking Free or Total Chlorine measurements, several basic rules should be followed:
• Never use cuvettes with scratches or cracks because they can cause inaccurate readings.
• Always cap the cuvettes to avoid spillage of the sample into the instrument.
• Always close the lid of the instrument during measurement.
• If is possible use the same cuvette for zero and sample reading.