To restore the factory calibration for the
currently selected range, press

CAL CHECK

key when in the main screen. The first screen
for the GLP is displayed. Press the

Delete

key to initiate the calibration delete
procedure and then press the

CFM key to

delete the user calibration and restore the
Factory calibration.
Note: Only the user calibration for the

current selected range is deleted.

CALIBRATION ERROR MESSAGES

• The calibration is successfully performed

if the reading is in certain limits. If the
CAL CHECK™ standard value is too high,
the display will show a high standard
message. If this message appears, check
if the correct cuvette was used.

• If the CAL CHECK™ standard value is too

low, the display will show low standard
message. If this message appears, check
if the correct cuvette was used.

RESTORE FACTORY CALIBRATION