beautypg.com

Hanna Instruments HI 98150N User Manual

Page 39

background image

39

pH CLEANING PROCEDURE

• General

Soak in Hanna

HI 7061 or HI 8061 General

Cleaning Solution for approximately ½ hour.

• Protein

Soak in Hanna

HI 7073 or HI 8073 Protein

Cleaning Solution for 15 minutes.

• Inorganic

Soak in Hanna

HI 7074 Inorganic Cleaning

Solution for 15 minutes.

• Oil/grease

Rinse with Hanna

HI 7077 or HI 8077 Oil

and Fat Cleaning Solution.

IMPORTANT: After performing any of the cleaning procedures, rinse the
electrode thoroughly with distilled water and soak the electrode in
HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking
measurements.