pH CLEANING PROCEDURE
• General
Soak in Hanna
HI 7061 or HI 8061 General
Cleaning Solution for approximately ½ hour.
• Protein
Soak in Hanna
HI 7073 or HI 8073 Protein
Cleaning Solution for 15 minutes.
• Inorganic
Soak in Hanna
HI 7074 Inorganic Cleaning
Solution for 15 minutes.
• Oil/grease
Rinse with Hanna
HI 7077 or HI 8077 Oil
and Fat Cleaning Solution.
IMPORTANT: After performing any of the cleaning procedures, rinse the
electrode thoroughly with distilled water and soak the electrode in
HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking
measurements.