Hanna Instruments HI 98150N User Manual
Page 21
21
S E T U P
S E T U P
S E T U P
S E T U P
S E T U P
Setup mode allows viewing and configuring the following parameters:
• Calibration Alarm Time Out (pH range only)(dAY)
• Custom buffers (1 or 2) (pH range only) (Custom)
• One-point calibration behavior (pH range only) (1Pnt)
• Current Time (hour & minute)
• Current Date (month, day & year)
• Beep Status (bEEP)
• Instrument ID (InSld)
• Auto-off backlight timer (LIGH)
• Auto power off timer (AOFF)
• Temperature Unit
To enter SETUP mode, press
Setup button directly on the HI 98140 and
2nd then Setup on the HI 98150 & HI 98160 while the instrument is
in measurement mode.
Select the desired setup parameter using the
ARROW keys.
Press
CAL if you want to change the item
value. “
CFM” tag and the selected item (e.g.
hour, in setting up the correct time) will start
blinking.
Press the
ARROW keys to change the displayed value.
If there is another item to be set (e.g. minutes), press
Range (HI 98150
&
HI 98160), 2nd then Mode (HI 98140). The other item will start
blinking.
Press the
ARROW keys to change the
displayed value.
Press
CFM to confirm or CAL to escape
without changing.
Press the
ARROW keys to select the next/
previous parameter.
Press
Setup to exit SETUP menu at any time.
The following table (see page 22) lists the SETUP parameters, their valid
values range and the factory settings (default). Use to move through
parameters in this order.