29
CHR xx
Change the instrument range according with the parameter
value (xx):
• xx=01 pH range/0.01 resolution
• xx=02 pH range/0.1 resolution
• xx=03 mV range
• xx=04 Relative mV range
The instrument will answer for these commands with:
> where: COMMANDS REQUIRING AN ANSWER The instrument will answer for these commands with: where the checksum is the bytes sum of the answer string sent as 2 ASCII Causes the instrument to send a complete set of readings in • pH and temperature reading ( HI 98140), and mV reading ( HI 98150 & HI 98160) on pH range. • Rel mV, absolute mV and temperature reading on Rel mV range ( HI 98150 & HI 98160). The answer string contains: • Meter mode (2 chars): • 01 - pH range (0.01 resolution) • Meter status (2 chars of status byte): represents a 8 bit hexadecimal encoding.
characters.
All the answer messages are with ASCII characters.
RAS
according with the current range:
• 02 - pH range (0.1 resolution)
• 03 - mV range
• 04 - Rel mV range
• 0x10 - temperature probe is connected
• 0x01 - new GLP data available
• 0x02 - new SETUP parameter