To enter Rel mV operation press

2nd and Mode from mV measurement.

• Place the ORP electrode into solution or

standard.

• Wait for a stable mV to be displayed.
• Press

CAL when the instrument is in

RELATIVE mV measurement mode. The
relative mV value is displayed on the
primary LCD and the absolute mV value
on the secondary LCD.

• Use the

ARROW keys to change the displayed relative mV value.

• When the reading is stable, in mV range and the Relative mV offset

is inside the offset window (±2000 mV), “

CFM” tag blinks.

• Press

CFM to confirm the relative mV calibration. The instrument will

return to measurement mode.

• If the absolute mV reading is out of range or the Relative mV offset

is out of the offset window, “

WRONG” tag will blink. Verify sensor tip

is fully submersed in solution and electrode is fully plugged in.