beautypg.com

Hanna Instruments HI 98150N User Manual

Page 12

background image

12

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, “

CFM”

tag blinks.

• Press

CFM to confirm calibration.

• The calibrated value is then displayed on

the primary LCD and the secondary LCD
will display the second expected buffer
value, together with “

CAL” and “Cal

Point 2” tags.

• After the first calibration point is confirmed,

submerse the pH electrode and the temperature
probe (if required) approximately 3 cm (1¼”) into the second buffer solution
and stir gently. The temperature probe should be close to the pH electrode.

• If necessary, press the

ARROW keys to select the appropriate buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, “

CFM”

tag blinks.

• Press

CFM to confirm calibration.

• The calibrated value is then displayed on the primary LCD and the

secondary LCD will display the third expected buffer value.

• After the second calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”) into
the third buffer solution and stir gently. The temperature probe
should be close to the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, “

CFM”

tag blinks.

• Press

CFM to confirm calibration.

• The calibrated value is then displayed on the primary LCD and the

secondary LCD will display the fourth expected buffer value.

• After the third calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”) into
the fourth buffer solution and stir gently. The temperature probe
should be close to the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, “

CFM”

tag blinks.

• Press

CFM to confirm calibration.

This manual is related to the following products: