• The calibrated value is then displayed on the primary LCD and the

secondary LCD will display the fifth expected buffer value.

• After the fourth calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”) into
the fifth buffer solution and stir gently. The temperature probe
should be close to the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, “

CFM”

tag blinks.

• Press

CFM to confirm calibration.

• The instrument stores the calibration values and

returns to normal measurement mode.

Note: For HI 98140 calibration proceed as in the FIVE-POINT CALIBRATION

above. After the third point is confirmed the instrument stores the
calibration value and returns to measurement mode.

FOUR, THREE or TWO-POINT CALIBRATION
• Proceed as described in FIVE-POINT CALIBRATION section.
• Press

CAL after the appropriate accepted calibration

point. The instruments will return to measurement
mode and will memorize the calibration data.

ONE-POINT CALIBRATION
Two SETUP selectable options are available for one-point calibration:

Pnt” and “OFFS”. See page 22 for Setup information using “1Pnt”

parameter.
If the “

Pnt” option is selected, the adjacent slopes will be recalculated

using the new calibration value.
If the “

OFFS” option is selected, an electrode offset correction is performed

keeping the existing slopes unchanged, but offsetting all by same value.
• Proceed as described in FIVE-POINT CALIBRATION section.
• Press

CAL after the first calibration point was confirmed. The

instruments will memorize the one-point calibration
data and will return to measurement mode.

Notes: • Press Range to toggle between pH buffer

and temperature reading during calibration.

• If the value measured by the instrument is not close to the

selected buffer, “

WRONG” “ ” and “WRONG” “ ” tags

will blink alternately. In this case check if the correct buffer
has been selected or used, or regenerate the electrode by
following the Cleaning Procedure (see page 39). If necessary,
change the buffer or the electrode.

• If the buffer temperature or the manual temperature exceeds

the temperature limits of the buffer, “

WRONG” “ ” tags

and temperature reading will blink.