• the calibration time, yymmddhhmmss (12 chars).

• ISE calibration data (if available), which contains:

• the number of calibrated standards (1 char)
• the ion charge, with sign (2 chars)
• the calibration slope, with sign and decimal point (7 chars)
• the calibration time, yymmddhhmmss (12 chars)
• standards information (for each standard)

• type (1 char): 0 - always standard solution.
• status (1 char):N (new) - calibrated in last calibration;

O (old) - from an old calibration.

• warnings during calibration (2 chars): 00 - no warning.
• standard value, with sign and decimal point (7 chars).
• calibration time, yymmddhhmmss (12 chars).

PAR

Requests the setup parameters setting.

The answer string contains:

• Instrument ID (4 chars)
• Calibration alarm time out (2 chars)
• SETUP information (2 chars): 8 bit hexadecimal encoding.

• 0x01 - beep ON (else OFF)
• 0x04 - degrees Celsius (else degrees Fahrenheit)
• 0x08 - Offset calibration (else Point calibration)

• Auto-off/Light time (3 chars)
• Auto power off time (3 chars)
• The number of custom buffers (1 char)
• The custom buffer values, with sign and decimal point,

for each defined custom buffer (7 chars)

• The ion charge (2 chars)

NSLx

Requests the number of logged samples (4 chars).

The command parameter (1 char):

• P - request for pH range
• M - request for mV and Rel mV ranges
• I - request for ISE range

LODPxxx

Requests the xxx

th

pH record logged data.

LODMxxx Requests the xxx

th

mV/Rel mV record logged data.

LODIxxx

Requests the xxx

th

ISE record logged data (HI 98172).

LODPALL Requests all pH Log on demand.
LODMALL Requests all mV/Rel mV Log on demand.