• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the

selected buffer, “CFM” tag blinks.

• Press CFM to confirm calibration.
• The calibrated value is then displayed on the

primary LCD and the secondary LCD will display
the second expected buffer value, together with
“CAL” and “Cal Point 2” tags.

• After the first calibration point is confirmed, immerse the pH electrode and the

temperature probe approximately 4 cm (1½”) into the second buffer solution
and stir gently. The temperature probe should be close to the pH electrode.

• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, “CFM” tag blinks.
• Press CFM to confirm calibration.
• The calibrated value is then displayed on the

primary LCD and the secondary LCD will display
the third expected buffer value.

• After the second calibration point is confirmed, immerse the pH electrode

and the temperature probe approximately 4 cm (1½”) into the third
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.

• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, “CFM” tag blinks.
• Press CFM to confirm calibration.
• The calibrated value is then displayed on the

primary LCD and the secondary LCD will display
the fourth expected buffer value.

• After the third calibration point is confirmed, immerse the pH electrode

and the temperature probe approximately 4 cm (1½”) into the fourth
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.

• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, “CFM” tag blinks.
• Press CFM to confirm calibration.
• The calibrated value is then displayed on the

primary LCD and the secondary LCD will display
the fifth expected buffer value.

LOG CFM

LOG CFM

LOG CFM

LOG CFM