Hanna Instruments HI 98170 User Manual
Page 33

3 3
CHR xx
Change the instrument range according with the parameter
value (xx):
• xx=01 pH range/0.01 resolution
• xx=02 pH range/0.1 resolution (HI 98170, HI 98171)
• xx=03 mV range
• xx=04 Relative mV range
• xx=05 ISE range (HI 98172)
The instrument will answer for these commands with:
> where: COMMANDS REQUIRING AN ANSWER The instrument will answer for these commands with: where the checksum is the bytes sum of the answer string sent as 2 ASCII characters. Causes the instrument to send a complete set of readings in • pH and temperature reading (HI 98170), and mV reading (HI 98171) on pH range. • Rel mV, absolute mV and temperature reading on Rel mV range (HI 98171, HI 98172). • ppm, mV and temperature reading on ppm range (HI 98172). The answer string contains: • Meter mode (2 chars): • 01 - pH range (0.01 resolution) • Meter status (2 chars of status byte): represents a 8 bit hexadecimal encoding.
All the answer messages are with ASCII characters.
RAS
according with the current range:
• 02 - pH range (0.1 resolution)
• 03 - mV range
• 04 - Rel mV range
• 05 - ISE range