Hanna Instruments HI 98170 User Manual

Page 20

background image

2 0

• The calibrated value will be displayed on the

primary LCD and the second expected standard
value on the secondary LCD, together with “CAL”
and “Cal Point 2” tags.

Note: The instrument will automatically skip the standard solution used for

the first point.

• After the first calibration point is confirmed, immerse the ISE electrode

approximately 4 cm (1½”) into the second calibration solution.

• If necessary, press the ARROW keys to select a different standard value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the “CFM”

tag blinks.

• Press CFM to confirm calibration.
• The calibrated value will be displayed on the

primary LCD and the third expected standard
value on the secondary LCD.

• After the second calibration point is confirmed, immerse the ISE electrode

approximately 4 cm (1½”) into the third calibration solution.

• If necessary, press the ARROW keys to select a different standard value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the “CFM”

tag blinks.

• Press CFM to confirm calibration.
• The calibrated value will be displayed on the

primary LCD and the fourth expected standard
value on the secondary LCD.

• After the third calibration point is confirmed, immerse the ISE electrode

approximately 4 cm (1½”) into the fourth calibration solution.

• If necessary, press the ARROW keys to select a different standard value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the “CFM”

tag blinks.

• Press CFM to confirm calibration.
• The calibrated value will be displayed on the

primary LCD and the fifth expected standard
value on the secondary LCD.

• After the fourth calibration point is confirmed, immerse the ISE electrode

approximately 4 cm (1½”) into the fifth calibration solution.

• If necessary, press the ARROW keys to select a different standard value.
• The “ ” tag will blink on the LCD until the reading is stable.

LOG CFM

LOG CFM

LOG CFM