Troubleshooting

68

FTB-5700

Solving Common Problems

Unable to find a reflective

fiber end. Please verify that

the fiber under test is

properly connected and

terminated by a UPC

connector.

Too much loss in the fiber

under test

³

Clean the fiber end.

³

Add a reflective

termination at the end

of the fiber.

Optical power is too low.

Too much loss in the fiber

under test.

³

Clean the fiber end.

³

Add a reflective

termination at the end

of the fiber.

Unable to find a valid

wavelength range.

³

Too much loss in the

fiber under test.

³

The wavelength range is

too large.

³

Clean the fiber end.

³

Add a reflective

termination at the end

of the fiber.

³

Reduce the wavelength

range in the Acquisition

tab of the Setup

window, or select

Default as the

wavelength range value.

I cannot open my reports.

Some of the report files are

missing.

Make sure that you have

kept all of the report’s

support files with the report

file itself.

Message

Possible cause