ISA-60M

ENMET Corporation

12

5.0 Maintenance

The

ISA-60M

has no specific preventative maintenance requirements. Entering the maintenance menu, as outlined in Section

5.2 may change instrument configurations.

The

MRI-5175

sensor will require periodic calibration and replacement. Oxygen sensor calibration should be preformed on at

least a quarterly basis. The sensor has an estimated lifetime of 5 years. The sensor should be replaced when it will not calibrate
or shortly before the 5 year time frame, to ensure continuous instrument operation. Sensor calibration instructions can be found
in Section 6.0.

5.1 Cleaning Instructions

C

AUTION

:

Never spray a cleaning solution on the surfaces of the

ISA-60M

or

MRI-5175

devices.

Clean the exterior of the

ISA-60M

and

MRI-5175

enclosures with a mild soap solution on a clean, damp cloth. Do not soak

the cloth with solution so that moisture drips onto, or lingers on, external surfaces.

Under no circumstances should organic solvents such as paint thinner be used to clean instrument surfaces.

5.2 Maintenance Menu ISA-60M

The maintenance menu diagram is shown in Figure 8 Maintenance Menu Flow Chart. From the operational display, press
the

OPTION

switch 4 times; "

Enter

MAINTENANCE Menu

" is displayed.

Figure 8: ISA-60M Maintenance Menu Flow Chart.

O = Press Option
S = Press Select

!