beautypg.com

3 power supply, 5 initial calibration – ENMET ISA-60M with MRI-5175 User Manual

Page 12

background image

ISA-60M

ENMET Corporation

9

Figure 5: ISA-60M Relay and Input and Output Terminals

3.3 Power Supply

The input power can vary from 100 to 240

V

AC

, 50/60 Hz. Mains power should be connected to the Power Input Terminal J23

and the ground screw J21. See Figure 5 for location.

W

ARNING

:

Continuous gas detection and alarm systems become inoperative upon loss of primary power.

Upon supplying power to the

ISA-60M

:

The green power on LED is lit.

The display backlight is lit, and instrument will step through a start-up sequence: unit serial number, software revision
and gases monitored may be shown on the display.

The instrument may go into alarm briefly, but the sensors stabilize quickly. If the instrument persists in alarm, acknowledge the
alarm by pressing the

A

UDIO

D

EFEAT

/A

LARM

A

CKNOWLEDGE

switch. If alarm persists longer than 30 minutes, call

ENMET

customer service personnel.

Mains power line is fused for power supply protection. Fuse is 5

x

20mm, 0.630Amp is located in

FH2

, see Figure 5

3.5 Initial Calibration

All instruments are calibrated at the factory. You may, if a calibration kit is available, calibrate the

MRI-5175

3 – 4 hours after installation. See Section 5.0, Maintenance, for calibration instructions.

!

Relay 1
Channel 1
Alarm 1

Relay 2
Channel 2
Alarm 1

Relay 3
Channel 3
Alarm 1

Relay 4
Channel 4
Alarm 1

Relay 5
Channel 1-4
Alarm 2

Relay 6
Ch 1-4 / System
Fault

J16 Detector Primary

Channel 1

J18 Connector

4-20mA Input

Channel 2

J19 Connector

4-20mA Input

Channel 3

Connector 2 (J4)
Channel 3 & 4
4-20mA Output

Connector 1 (J7)
Channel 1 & 2
4-20mA Output

Connector RS485

Connector RS232

Connector RS485

Power Input, Terminal J23

L-Line
N-Neutral

Ground Screw J21

Fuse Holder FH2