OMNI-4000 S/N 6705 and above

44

Inputting the Calibration Gas Value
This procedure defines the content of the calibration gas to be applied to the explosive gas channel sensor. The
maximum permissible content is 100% LEL.
The display reads:

To modify this value:

• Select the units by repeatedly pressing the MENU switch. When the desired value appears, press BACKLIGHT

switch.

• Select the tenths of a unit by repeatedly pressing the MENU switch. When the desired value appears, press

BACKLIGHT switch.

• Select the hundredths of a unit by repeatedly pressing the MENU switch. The units and tenths can be accessed

again by pressing the BACKLIGHT switch. When the calibration gas value is complete, press the ENTER
switch.

The value of the calibration gas to be applied is memorized.
Calibrating the Zero Point
This must be performed in an area that is free of gas contamination. The display reads, for instance:

• Press the ENTER switch. The zero is set automatically by the OMNI-4000 electronics for future internal

calculations.

Calibrating the Sensitivity
The display shows the offset value obtained during the zero calibration phase, for example:

The value is always stated in % LEL CH

4

. Do not press a switch.

• Apply the calibration gas. You must be able to just hear the gas. The flow rate must be approximately 1 l/min.
• When the readings have stabilized, press the ENTER switch. The display shows:

Press: • The ENTER switch to confirm the reading. The return to the initial "Calibrate a Sensor" menu is automatic

unless there is a problem (See "Problems" paragraph). It is now possible to calibrate another channel by
pressing ENTER. It should be noted that the channel has automatically been incremented. If there are not
further channels to calibrate, go to the "Return to The Operator Mode" paragraph.

• Pressing any other switch results in the return to the initial "Calibrate a Sensor" menu without memorizing the

value entered during the above calibration procedure.

Cal gas

100%LEL

CH4

Zero value

001 LEL

CH4

Span value

001 LEL

CH4

Accept:

ENTER

Cancel: ¨