OMNI-4000 S/N 6705 and above
ENMET Corporation
25
The action to be taken depends upon the MESSAGE displayed.
MESSAGE: Battery low
Recharge the batteries
There is approximately 20 minutes of operating time left, after which the device
stops operating.
This alarm cannot be acknowledged; pressing ENTER has no effect.
MESSAGE:
>100LEL
Only concerns the explosive gas channel
Effect:
Latching of the display quadrant concerned
Continuous audible alarm which cannot be acknowledged
Explosive gas channel and general alarm indicator lamps are lit
Return to normal operation by turning OFF and restarting the OMNI-4000
Proceed with care in the area as there is a concentration of explosive gas above the
LEL
N
OTE
: This alarm occurs only when the "0-100% GAS" option has not been selected. If the OMNI-4000 is
programmed with this option, the reading goes directly to the "0-100% GAS" range; an instrument programmed with
this option must include an oxygen sensor to be operational.
MESSAGE:
Over range
Only concerns the toxic gas channels
Effect:
Continuous audible alarm which cannot be acknowledged
Toxic channel and general alarm indicator lamps are lit
Leave the area immediately as there is an excessive concentration of toxic gas
MESSAGE:
New Cal
Automatic zero reset (auto set) cannot be performed because, for instance, there is
an excessive drift from zero of a sensor
Replace the affected sensor. See Section 3.