Calibration test report using the internal weight
Key operation
Step 1 Press and hold the CAL key until displaying
[ [
, Release the key.
Step 2 The balance displays T | and performs the calibration test automatically.
Step 3 The zero point is measured and this value is displayed.
Step 4 The internal weight is measured and this value is displayed.
Step 5 If the calibration test report is output ,
[ , j P
is displayed and the GLP
data is output.
Step 6 The balance returns to the weighing mode automatically.
Command operation
Step
1
Transmit the
TST
command to the balance.
Step 2 The balance performs the calibration test automatically.
Step 3 If the callbr'-tion test report is output, the GLP data is output.
Step 4 The balance returns to the weighing mode automatically.
AD-8121 format
Data format
III' LI I
a
K: [>
(5 i?. . .
£; H
12; ::5 ;5 2.7^^ 77
i: i)
77 77 77 4- / iZi 77
775;: 21 4 2 PM
Cii:. . . . . . T E S T l I N I „ >
77 C T 77 77 7..
77;. ;7;; ;77 ;77; ;Z; 77;
77 ;7; ;7; = ;7; 77; ;7; 77: 77;
77;
:2 I ;3 77 77 T 77 7777:
— Factory —
—
Product —
Serial number
— ID number -
----- Date ---------
----- Time --------
— Cal. test type
- Zero point value
Target mass value-
Target weight-
Signature
^ Space mark, ASCII 20H.
cTERM>
Terminator mark, ‘-
f
or ^
r
.
C
r
Carrige return mark, ASCII ODH
>-F Line feed mark, ASCII OAH
I I I I U L
A &
D
MODEL
GR-Z00
S/N
12345678
ID
ABODE FGH
DM i
TIME
CAL .TESTCINT. )
ACTUAL
0.0000
Q
+ 200.0002
g
TARGET
+ 200.0000
g
SIGNATURE
----------------------------
^TERM>
Example of GR-200
10, The ID Number, GLP Report
Page 42
Functions