Calibration test – A&D GR Series GR-300 User Manual
Page 25
Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Calibration Test
This function tests the balance for the accuracy using an external weight.
A weight which is used for the calibration test is called "the target weight".
The weight which you have is called "the external weight".
Operation
Step 1 Connect the AC adapter and warm up the balance
for at least one hour with nothing on the pan.
Step 2 Press and hold the CAL key until displaying
and then release the key.
LI U L
Press and hold the key
n
u
Step 3 The balance displays
□
If you want to change the target weight value,
proceed to step 4.
□
If you use the stored target weight value in the
balance, proceed to step 5.
Step 4 Press the j RANGE key and adjust the calibration
weight value using the following keys.
RE-ZERO [ key
RANGE key
PRINT key
CAL]key
The key to set the value of the
digit selected.
Thekeytoselectthedigittochange
value.
The key to store a new weight
value and return to step 3.
The key to cancel this change
and return to step 3.
Product
Usable weight
Adjustable range
GR-120
100g, 50g
GR-200
200g, 100g
-1-15.9 mg ~ -15.0 mg
GR-300
200g, 300g
GR-202
200g, 100g
Note
Digits cyclically change using the , RE-ZERO | key.
ex. 0mg^-h15mg-^-15mg-^0mg
Step 5 Confirm that there is nothing on the pan and press
the PRINT key. The balance measures the zero-
point and displays it. Prevent vibration and drafts
from affecting the balance.
7. Calibration
Page 24
ciuc
-Release the key
,Viod
ex. target weight
value = 200.0017g
Initial target weight
value is 200.0000g
at GR-200.
r r
n
L L
LI
o
rinnnn
u.uuuu
g
(To Next Page)
Adapting to environnnent