Calibration using the internal weight – A&D GR Series GR-300 User Manual
Page 21
Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Calibration Using the Internal Weight
This function calibrates the balance using the internal weight.
Operation
Step 1 Connect the AC adapter and warm up the balance for
at least one hour with nothing on the weighing pan.
Step 2 Press the CAL| key to start calibration.
Step 3 The balance displays
\ [ n j ~ m \
and performs calibra
tion. Prevent vibration and drafts from affecting the
balance.
Step 4 Ifthe "CLP output ( ,n/-o)" of the "Function Table" isset
to
232C interface.
or "P", "Calibration Report" is output from RS-
Step 5 The balance will automatically return to the weighing
mode after calibration.
Step 6 Test the accuracy of weighing using the calibration test
function or by using a certified test weight.
Control of this Calibration
Calibration using the internal weight is controlled by a
parameter in "Permission or prohibition". Refer to section
"8, Function Switch and Initialization".
n n M ( i
II U M M
----------- A--
-Calibration Using the internal Mass
0 ; Not used
1 :
Used