6. Response Adjustfinent
This function stabilizes the weight value, reducing the influence on weigh
ing that is caused by drafts and/or vibration at the place where the balance
is installed. This function adjusts by automatically analyzing the environ
ment or by hand-operation. The function has three stages as follows :
Indicator
Function table
Summaries
FAST
^ - J
L Ü n U U
Fast response. Sensitive value
MID.
r _ 1 1
L U > 1 LI t
SLOW
r . ^ J Zl
L Û n U L
Slow response. Stable value
Response Indicator
RESPONSEÎrASTIIMlD.lfSlLÜWl'
fin 11 11 ft
yyijyij
Note
□
If the automatic response adjustment is to awkward, Try to refine it using
the section "Manual Response Adjustment".
□
The response adjustment can be changed at "Condition {[and)" of
"Environment & Display {bnSi-nc)" in the function table. Refer to "9.
Function table".
Automatic Response Adjustment
This way automatically updates the response adjustment by analyzing the
influence of the environment on the weight data.
Operation
Step 1 Press and hold the
[MODE
| key until
[RESPONSE
displayed.
is
Press and hold the key.
Step 2 The balance analyzes the influence and updates the
response adjustment. If you want to cancel this
update, press the |CAL| key.
Caution
Do not allow vibration or drafts to affect the balance.
Step 3 The balance returns to the weighing mode automati
cally and displays the updated response indicator.
RESPONSE
- Release the key.
RESPONSE
r U
L ) I
... ,,, ...
\
}
RESPONSE
tvî5~l
$
u 1)}
$ $
i W
y.uuuu
s
6. Response Adjustment