Accu-Chek Aviva Combo User Manual
Page 243
241
8
Display Shows Possible Cause(s)
Possible Solution(s)
InvalidTarget
Range
j
Thelowertargetrangevalue
isabovetheuppertarget
rangevalue.
j
Resetthetargetrangevalues.
j
Enterthecorrectlowertargetrangevalueand
uppertargetrangevalue.SeeChapter5,“Changing
MeterSettings.”
j
Therangeselectedconflicts
withyourhyperand/orhypo
warninglimitsettings.
j
Resetrangeorrevisewarninglimitsettingsand
re-enterrange.
j
Enterthecorrectlowertargetrangevalueand
uppertargetrangevalue.SeeChapter5,“Changing
MeterSettings.”
LObGWarning
j
Yourbloodglucosemaybe
lowerthanthemeasuring
rangeofthesystem.
j
Treatyourlowbloodglucoseasrecommendedby
yourhealthcareprofessional.
j
Ifyouareexperiencinganyofthecommon
symptomsoflowbloodglucose,contactyour
healthcareprofessionalimmediately.
MeterandPump
NotPaired
j
Meterandpumphavenot
beenpaired.Youcannotuse
thepumpfunctionswithout
pairingthemeterandthe
pump.
j
Youmustpairthemeterandpumptousethese
features.
j
Forinstructionsonhowtopairthemeterandpump,
seetheAdvancedOwner’sBooklet.
MeterBatteryLow
j
Batterypowerislow.
j
Installnewbatteries.SeeChapter7,“Careand
Maintenance.”