8.2 Troubleshooting the Accu-Chek Aviva Combo Meter

Display Shows Possible Cause(s)

Possible Solution(s)

The­display­is­blank­
or­the­meter­will­
not­turn­on.

j

­­Batteries­are­dead.

j

­­Install­new­batteries.­See­Chapter­7,­“Care­and­

Maintenance.”

j

­­Display­is­damaged.

j

­­Contact­Roche.

j

­­Meter­is­defective.

j

­­Contact­Roche.

j

­­Extreme­temperature­–­the­

temperature­is­outside­the­
meter­operating­range.

j

­­Move­the­meter­to­an­area­with­proper­temperature.­

Wait­five­minutes­before­turning­on­the­meter.­Do­
not­artificially­heat­or­cool­the­meter.

Bluetooth­icon­
flashing

j

­­Communication­ended­due­to­

a­button­press­on­the­pump.

j

­­Check­the­pump­and­continue­operation­of­the­

pump­manually.

j

­­Ensure­the­pump­has­Bluetooth­wireless­technology­

turned­on.

j

­­Meter­and­pump­are­outside­

of­communication­range.

j

­­Ensure­pump­is­within­communication­range.