232
Display Shows Possible Cause(s)
Possible Solution(s)
BolusTooHigh
j
Thebolusisabovethe
allowedamount.Thebolus
wassetto50Units.
j
Checktheaccuracyofallentries.Ifnecessary,
contactyourhealthcareprofessional.
BolusTooLow
j
Theinsulinpumpcannot
delivera0.1Unitofmultiwave
bolus.
j
Thepumpwilladjustthebolusto0.2Units.
CarbohydrateRatio
j
Carbohydrateratioisoutside
oftheacceptablemeter
range.
j
Checkyourentriesandcontactyourhealthcare
professionaltodeterminetheappropriatesettings.
CodeKeyMissing
j
Theactivationchipisnot
inserted.
j
Turnthemeteroffandinserttheactivationchip.If
youneedanactivationchip,pleasecontactRoche.
Communication
Lost
j
Pumpisoutofrangeofthe
meter.
j
Bolusdeliverycontinuing,seepumpscreen.
j
EnsurethemeterandpumphaveBluetoothwireless
technologyturnedon.
j
Ensurethemeterandpumparewithin
communicationrange.
j
Ifthelowbatteryiconisdisplayedonthemeter,the
metercannotcommunicatewiththepump.Replace
thebatteries.
j
Usethepumptomonitororcancelaboluswhichis
intheprocessofbeingdelivered.