Display Shows Possible Cause(s)

Possible Solution(s)

Invalid­Hyper­
Values

j

­­The­hyper­warning­limit­value­

must­be­greater­than­all­of­
your­target­ranges­in­the­time­
block­settings.

j

­­Reset­the­hyper­warning­limit­or­revise­the­target­

ranges­in­time­blocks­and­re-enter­the­hyper­
warning­limit.

j

­­Enter­a­hyper­warning­limit­that­is­above­the­target­

ranges­of­your­time­blocks.­See­Chapter­5,­
“Changing­Meter­Settings.”

Invalid­Hypo­Values

j

­­The­hypo­warning­limit­value­

must­be­less­than­all­of­your­
target­ranges­in­the­time­
block­settings.

j

­­Reset­the­hypo­warning­limit­or­revise­the­target­

ranges­in­time­blocks­and­re-enter­the­hypo­
warning­limit.

j

­­Enter­a­hypo­warning­limit­that­is­below­the­target­

ranges­of­your­time­blocks.­See­Chapter­5,­
“Changing­Meter­Settings.”

Invalid­PIN

j

­­The­wrong­PIN­was­entered.

j

­­Select­OK­and­re-enter­the­PIN­shown­on­the­pump­

display.

Invalid­Record­
Time/Date

j

­­The­time/date­entered­is­

invalid­(Add­Data­entries­
cannot­be­set­to­occur­in­the­
future).

j

­­Re-enter­time/date.