Accu-Chek Aviva Combo User Manual
Page 241
239
8
Display Shows Possible Cause(s)
Possible Solution(s)
InfrequentPump
Communication
j
Atleast2weekshavepassed
sincethemeterandpump
havecommunicated.
j
EnsurethemeterandpumphaveBluetoothwireless
technologyturnedon.
j
Ensurethemeterandpumparewithin
communicationrange.
j
Ifthelowbatteryiconisdisplayedonthemeter,the
metercannotcommunicatewiththepump.Replace
thebatteries.
j
ItisimportanttouseBluetoothwirelesstechnology
communicationbetweenthemeterandpump
regularlyifyouutilizebolusadvice.
InsulinSensitivity
j
Insulinsensitivityisoutsideof
theacceptablemeterrange.
j
Checkyourentriesandcontactyourhealthcare
professionaltodeterminetheappropriatesettings.
InvalidBolus
AdviceTimes
j
Theactingtimevalueisless
thantheoffsettime.
j
Theactingtimevaluemustbesetequaltoor
greaterthantheoffsettime.Resettheactingtime
valueorrevisetheoffsettime.
InvalidDate
j
Thedateenteredisinvalid
(datesforreminderscannot
besettooccurinthepast).
j
Re-enterdate.