YSI 610DM User Manual
Page 16

YSI Incorporated
Displays and Loggers
14
SECTION 6 SAMPLING
Sampling applications refer to readings that are taken while the user is present and
controlling the collection of data.
In sampling applications the total data collection time is relatively brief, from a few
minutes to perhaps a few hours. Examples: taking one set of readings at the outfall of a
wastewater treatment plant once a day, taking readings at one mile intervals along the
length of a river, checking the DO concentration once a week in several ponds, or
spending an entire day profiling a lake.
The YSI 610 has several powerful features that make data collection easy during
applications. Because the interpretation and use of the Site List depends on which feature
you use, we recommend that you choose one method that best suits your needs, learn that
method and ignore the others. The following chart lists the major features of each type of
memory. The rest of this section explains in detail the uses of each type.
Stash
Simplest way to store readings
Data is not intended to be uploaded to a PC
Data is stored in 26 slots, each referred to by letter (A-Z)
Weather data can be stored in the same way
Add
Stores single sets of readings to an uploadable file
Only available on the YSI 610-DM
Capture
Stores several consecutive readings to an uploadable file
Only available on the YSI 610-DM.
Mark
Stores readings from many sites in a single, uploadable file
Only available on the YSI 610-DM.
6.1 STASH MEMORY
Stash memory is available on both the YSI 610-D and YSI 610-DM display/loggers.